Ellab TRACKSENSE PRO Data loggers

Each logger has multiple channels for recording data with a memory capacity of up to 60,000 data points. The state of the art technology allows for variable sample rates. A predefined start and stop time can be set and the sample rate can be adjusted within a specific temperature range.

The logger is made of AISI 316L stainless steel and the electronics are sealed in heat and moisture resistant material.

So Sánh

Mô tả

Ellab TRACKSENSE PRO Data loggers

Brand: Ellab – Denmark

TSP-sensorrange

Data Loggers

Each logger has multiple channels for recording data with a memory capacity of up to 60,000 data points. The state of the art technology allows for variable sample rates. A predefined start and stop time can be set and the sample rate can be adjusted within a specific temperature range.

The logger is made of AISI 316L stainless steel and the electronics are sealed in heat and moisture resistant material.

A Variety Of Loggers To Fit Your Application
The standard Pro logger is designed with interchangeable sensors and user replaceable batteries whereas the Mini and Micro are working with mounted sensors and user replaceable button cell batteries.

TrackSense® Pro Logger
TrackSense® Pro loggers are designed to be accurate and durable in the harshest conditions. All components have been selected and tested to withstand the high temperatures and pressures associated with steam sterilization.

loggerrange-v_Large
TrackSense® Pro X Logger
The TrackSense® Pro X logger is specially designed for -80°C applications such as lyophilization with low temperature and vacuum.

Sky_module_L

TrackSense® Pro XL Logger
The TrackSense® Pro XL is an extended logger with increased battery capacity. This logger is designed for the all applications in temperature range of -80 to +150˚C.

TrackSense® Pro Basic Logger
Many applications do not require the high temperature range supplied by TrackSense® Pro or Pro X loggers. The Basic logger is designed to be a more specific solution for applications below 105°C. Suitable applications would be ambient warehouse validations, pasteurization, humidity and temperature stability studies.

TSP_Pro_X_basic_L
In accordance with ITS 08 ATEXQ5928 TrackSense® Pro, Pro X and Pro Basic loggers, inclusive Sky and equipped with pressure sensors and relative humidity sensors can not be exposed to more than 105°C when working in an environment where intrinsically safe equipment is required.

TrackSense® Lab For Stability Studies
This logger is designed with an active LED light to indicate when a pre-programmed alarm situation has been triggered. The LED blinks green when the logger is running within the pre-programmed temperature band. If the measured temperature reaches outside of this band, the LED blinks red indicating an alarm situation.

TSP-LAB2_L
The TrackSense Lab logger is ideal for monitoring temperature or RH in long-term stability applications. Find further information about TrackSense Labhere.

TrackSense® Pro Mini And Micro
The Mini and Micro loggers are designed for applications where size is an issue. Each logger has a memory capacity of up to 14,500 data points. The low volume displacement makes these small loggers ideal for monitoring inside packaging. The TrackSense Pro Mini offers temperature monitoring only.

TSP_mini_micro_L
Find further information about the Mini loggers here. The TrackSense Pro Micro can be configured with temperature or pressure sensors. Find further information about the Micro loggers here.

Custom Fittings
Packing glands are available for inserting sensors into any variety of packaging. The glands are threaded to accept sensors and will maintain the seal when pressurized.

ungdung-logger

 
Logger

 

Dimension DxH (mm)

Data Storage

Battery Life Hours

Temperature Range °C

TrackSense® Pro X (ATEX certified)**

25 x 44

60,000

-80 to +150*

TrackSense® Pro XL  25 x 68.8  60,000  -80 to +150*

TrackSense® Pro  (ATEX certified)***

25 x 44

60,000

-50 to +150*

TrackSense® Pro Basic

25 x 44

60,000

-30 to +105 

TrackSense® Pro Mini

20 x 16

14,500

100

0 to +140

TrackSense® Pro Micro(Temperature)

15 x 22

14,500

75

-20 to +140 

TrackSense® Pro Micro
(Pressure)

15 x 40

7,200*

50

-20 to +140 

TrackSense® Pro Compact (Temperature)

25 x 35

14,500

-30 to +140 

TrackSense® Pro Compact X 25 x 35 14,500 -50 to +140
TrackSense® Pro Compact UX (Temperature) 25 x 60 14,500 -80 to +140
TrackSense® Pro Compact (Pressure)

25 x 55

7,200

-30 to +140 

TrackSense® Pro Lab

25 x 43.5

60,000

-30 to +100 

TrackSense® Pro Lab Mini

20 x 16

14,500

100

0 to +100

* Max 140 °C with Sky Module
** ATEX -60 to +105 °C
*** ATEX -30 to +105 °C

 

*Available data storage capacity is reduced from 14,500 to 7,200 due to internal temperature compensation of the TrackSense® Pro Micro logger pressure sensor.

The High Performance Battery

Battery life greatly depends on the application and logger use profile.
Extensive ramping and duration at higher temperatures can reduce the average life of the battery.
The battery incorporates the latest battery technology resulting in 2,000 hours of operation at a sample rate at 10 seconds at 121°C. For further information about battery life and sample rate,

Technical Specifications

 
Operating Logger: Lab, Compact, Compact X, Basic, Pro, Pro X 
Battery Type: Lithium
Capacity at 121 °C / 10 sec. Sample Rate:  
Temperature 1 Sensor 2,000 Hours
Temperature 2 Sensors

1,750 Hours

Temperature 4 Sensors (Lab QUAD) 1,250 Hours
Pressure 1,500 Hours
Pressure / Temperature 1,500 Hours
RH% 1,500 Hours
Conductivity 1,000 Hours
Vacuum    300 Hours
Intrinsically Safe: EEx ia IIC T3
Compliance: UL Listed

TRACKSENSE® PRO SENSOR

TrackSense® Pro Sensors
Ellab offers the largest range of different sensors with 1, 2 or 4 channels. Sensors are interchangeable, enabling the user to change sensors for different applications. This reduces costs because one set of TrackSense® Pro loggers can be used for temperature, humidity and pressure studies.

loggerrange-v_Large
The temperature sensors can be delivered in rigid, semi flexible and flexible material for ease of use. New sensors are available with or without LED that shows the status of the logger. An active, communicating logger is identified by a green blink.  Loss of communication between the logger and Access Point is signaled by a red blink. This feature makes it much easier to start larger groups of loggers and helps avoid using unprogrammed loggers. In the case of an LED sensor being used in combination with a Sky module, the LED will confirm communication status.

The TrackSense® Pro sensors cover parameters such as:

  • Temperature
  • Pressure
  • Relative Humidity
  • Rotation
  • AutoMarker
  • Conductivity

Temperature
The configuration of the sensors is also very extensive: 

  • Single Rigid Sensor
  • Double Rigid Sensor
  • Single Flexible Sensor
  • Double Flexible Sensor
  • Double Sensor With One Rigid And One Flexible Sensor
  • Quad Flexible Sensor


Extreme Temperature Sensors
The standard temperature range is 0°C to +150°C, but it is possible to ordersensors which can be calibrated and measure down to -196°C or up to +400°C.

The logger has to be placed outside of the environment when measuring at -196°C. When measuring from +150°C up to 400°C, an insulation pack (thermal barrier) is required. The principel is to insulate the logger for a limited amount of time keeping the logger below +150°C. Please consult your local Ellab representative for specific time and temperature limits.

Pressure

The pressure sensor can be used as a stand alone sensor to measure the pressure in e.g. the autoclave itself, but can also be mounted with a connection system to facilitate measurements in small containers without having to place the logger inside the container.

logger-sensor-pressure

Vacuum

The Vacuum sensor is ideal for lyophilization. Due to its “micro pirani” design this sensor offers extreme accuracy combined with trouble free placement in process either for batch monitoring or chamber validation.

TSP_Vacuum_XL

Relative Humidity

To achieve the best results in measuring relative humidity it is necessary to choose the right design of the sensor. The sensor needs to be protected if used in harsh, corrosive conditions such as an EtO sterilizer.
TSP-RH_L
If the sensor is not protected, initial results will be satisfactory but the conditions will affect the sensor and the sensor will eventually start drifting. Therefore, Ellab’s sensing element is protected by a filter but contains special openings to increase air run through.

AutoMarker
The AutoMarker sensor can be used to record events during a session. When started, time markers can be added to the session either by manual activation of the switch placed at the cable end or automatically when connected directly to a PLC using the PLC relay function.

auto_marker_1_L
Examples of AutoMarker sensor use are:

  • Logging of the holding time start/stop from autoclave controller
  • Logging of the door openings in warehouse storage, stability chambers, freezers, incubators etc.
  • Logging of general failure of critical components during process e.g. pump failure
  • Logging of emergency shelf cooling activation in freeze dryer

Conductivity
The Conductivity Sensor is compatible with TrackSense Pro Sky option to provide live data if required. Conductivity measurements apply to the control / monitoring of industrial processes and to validate the washing disinfecion processes in hospitals governed by regulation EN ISO 15883:2009.

Conductivity_large
Often conductivity is measured in combination with pH measurements.

Examples of applications the Conductivity Sensor would be used in are:

  • Pharmaceutical Industry
  • Hospitals
  • Fermentation
  • Food and Beverage
  • Sugar Industry
  • Petro-Chemisty
  • Chemical Industry
  • Power Plants

Technical Specifications for Rigid SS Standard Sensor

TSP_Rigid_SS_Standard_XL TSP_Rigid_SS_Standard

 
Type     Rigid SS Standard
Diameter     2 mm / 3 mm
Length     From 10 mm
Dimension of sensing element     1.2 x 1.6 mm
Position of measuring point:  
From round tip     3 mm
From sharp tip     6 mm
From conical tip     8 mm
Temperature measuring range     -196°C to +150°C
Accuracy:  
Low -196 to -80°C     ± 0.5°C
Low -80 to -50°C     ± 0.3°C
Low -50 to -40°C     ± 0.2°C
Low -40 to -25°C     ± 0.1°C
Low -25 to 0°C     ± 0.05°C
Standard 0 to 100 °C     ± 0.05°C
Standard +100 to +140°C     ± 0.05°C
Standard +140 to +150°C             ± 0.05°C
Sensor capacity     1 or 2     
Sensor response time:  
T-10%     0.11 sec. / 0.22 sec.
T-50%     0.35 sec. / 0.72 sec.
T-63%     0.5 sec. / 1.00 sec.
T-90%     1.10 sec. / 2.00 sec.
Pressure operating range     0.001 mBar to 10 Bar ABS 
Technical Specifications for Rugged Flexible Sensor
TSP_Rugged_Flex_XL TSP_Rugged_Flex
Type     Rugged Flexible                        
Diameter     4 mm
Length     500 mm / 1,000 mm                  
Dimension of sensing element     ø2.5 x 30 mm
Position of measuring point from tip:     5 mm
Temperature measuring range     -196°C to +140°C
Accuracy:  
Low -196 to -80°C     –
Low -80 to -50°C     –
Low -50 to -40°C     –
Low -40 to -25°C     –
Low -25 to 0°C     –
Standard 0 to 100 °C     –
Standard +100 to +140°C     –
Sensor capacity     1    
Sensor response time:  
T-63%     3.5 sec.
T-90%     7.0 sec.
Pressure operating range     –
Technical Specifications for Semi Flexible SS Standard
TSP_Semi_Flexible_SS_Standard_XL TSP_Semi_Flexible_SS_Standard
 
Type     Semi Flexible SS Standard      
Diameter     1.5 mm / 2 mm / 3 mm
Length     100 -1000 mm
Dimension of sensing element     0.8 x 3.0 mm/1.2 x 1.6 mm/
1.2 x 1.6 mm
Position of measuring point:  
From round tip     3 mm
From sharp tip     6 mm
From conical tip     8 mm
Temperature measuring range     -196°C to +150°C
Accuracy:  
Low -196 to -80°C     ± 0.5°C
Low -80 to -50°C     ± 0.3°C
Low -50 to -40°C     ± 0.2°C
Low -40 to -25°C     ± 0.1°C
Low -25 to 0°C     ± 0.1°C
Standard 0 to 100°C     ± 0.1°C
Standard +100 to +140°C     ± 0.1°C
Standard +140 to +150°C     ± 0.1°C
Sensor capacity     1 or 2
Sensor response time:  
T-10%     – / 0.11 sec. / 0.22 sec.
T-50%     – / 0.35 sec. / 0.72 sec.
T-63%     0.46 sec. / 0.41 sec. / 1.00 sec.
T-90%     0.67 sec. / 0.81 sec. / 2.00 sec.
Pressure operating range     0.001 mBar to 4 Bar ABS 
Technical Specifications for Flexible Low Sensors
TSP_Flexible_low_XL TSP_Flexible_low
Type     Flexible Low                            
Diameter     1.2 mm
Length     100 – 500 mm
Dimension of sensing element     0.8 x 3 mm
Position of measuring point:  
From round tip     3 mm
Temperature measuring range     -196°C to +100°C*
Accuracy:  
Low -196 to -80°C     ± 0.5°C
Low -80 to -50°C     ± 0.3°C
Low -50 to -40°C     ± 0.2°C
Low -40 to -25°C     ± 0.1°C
Low -25 to 0°C     ± 0.1°C
Standard 0 to 100°C     ± 0.1°C
Sensor capacity     1, 2 or 4
Sensor response time:  
T-63%     1.34 sec.
T-90%     2.55 sec.
Pressure operating range     0.001 mBar to 1 Bar ABS   
 * Can be sterilized at 125°C
Technical Specifications for Flexible Standard Sensors
TSP_Flexible_Standard_XLTSP_Flexible_Standard
Type     Flexible Standard                    
Diameter     1.8 mm
Length     100 – 1000 mm
Dimension of sensing element     1.2 x 1.6 mm
Position of measuring point:  
From round tip     3 mm
Temperature measuring range     -196°C to +140°C
Accuracy:  
Low -196 to -80°C     ± 0.5°C
Low -80 to -50°C     ± 0.3°C
Low -50 to -40°C     ± 0.2°C
Low -40 to -25°C     ± 0.1°C
Low -25 to 0°C     ± 0.1°C
Standard 0 to 100°C     ± 0.1°C
Standard +100 to +140°C     ± 0.1°C
Sensor capacity     1 or 2
Sensor response time:  
T-63%     2.16 sec.
T-90%     4.31 sec.
Pressure operating range     0.001 mBar to 4 Bar ABS   
Technical Specifications for Rigid Sensor High
TSP_Rigid_SS_high_XLTSP_Rigid_SS_high
Type     Rigid SS High       
Diameter     3 mm
Length     From 150 mm 
Dimension of sensing element     1.2 x 1.6 mm
Position of measuring point:  
From round tip     3 mm
From sharp tip     6 mm 
From conical tip     8 mm
Temperature measuring range     0 to +400°C
Accuracy:  
High 0 to +400°C     ± 0.5°C
Sensor capacity     1
Pressure operating range     0.001 mBar to 10 Bar ABS 
Technical Specifications for Stainless Steel sensors
Semi_flexible_SS_HighSemi_flexible_SS_High_Top
Type     Semi Flexible SS High             
Diameter     2.0 mm / 2.5 mm
Length     150 – 1000 mm
Dimension of sensing element     1.2 x 1.6 mm
Position of measuring point:  
From round tip     3 mm
From sharp tip     6 mm 
From conical tip     8 mm
Temperature measuring range     0 to +400°C
Accuracy:  
High 0 to +400°C     ± 0.5°C
Sensor capacity     1 or 2
Sensor response time:  
T-63%     1.46 sec. / 6.96 sec.
T-90%     2.87 sec. / 17.72 sec.
Pressure operating range     0.001 mBar to 4 Bar ABS   
 
Long term stability: The PT1000 sensing element will change less than 0.03% after 1,000 hours of usage at max temperature (600°C)
Calibration:                              Annual factory recalibration of sensors is recommended.
The ValSuite Software allows customers to do an offset calibration in the software

The Pressure Sensors

The pressure sensor can be used as a stand alone sensor to measure the pressure in e.g. the autoclave itself, but can also be mounted with a connection system to facilitate measurements in small containers without having to place the logger inside the container.

Technical Specifications
 
Material: Stainless Steel
Measuring Principal: Strain Gauge
Operating Temperature: 0 to +150°C
Operating Pressure: 10 mBar to 10 Bar Absolute
Measuring Range: 10 mBar to 6 Bar Absolute, Up To 7 Bar Can Be Ordered
Resolution: 1 mbar
Measuring Accuracy: ±0.25% Full Scale

The Combined Pressure/Temperature Sensor

The combined pressure/temperature sensor can be used as a stand alone sensor to measure pressure and temperature simultaneously in various types of applications, like e.g. a steam sterilizer.

Technical Specifications
TSP_kombitryk_L
  General Sensor Specifications
Material: Stainless Steel
Operating temperature: 0 to +150°C
Sampling: 1 Sec. To 24 Hours Auto Start/Stop
   
  Pressure Sensor Specifications
Measuring Principle: Strain Gauge
Operating Pressure: 10 mBar to 10 Bar Absolute
Measuring Range: 10 mBar to 6 Bar Absolute, Up To 7 Bar Can Be Ordered
Resolution: 1 mbar
Measuring Accuracy: ±0.25% Full Scale
   
  Temperature Sensor
Measuring Element: External PT1000
Length of Sensing Element: 20 mm
Measuring Range: 0 to +150°C
Resolution: 0.012°C
Measuring Accuracy: 0.05°C

 

Đánh giá

Chưa có đánh giá nào.

Hãy là người đầu tiên nhận xét “Ellab TRACKSENSE PRO Data loggers”